Apply the IQ reasoning
20 distinct application questions. This open-book self-study assessment exposes its answer key in the source and below. The suggested practice target is 85% (17/20), not a regulatory threshold, IQ acceptance rule or proof of independent competence. Unanswered questions must be completed before scoring. Answers are not sent to a server.
Answer key and explanations
1. The installed arrangement matches the reviewed scope — Limit the conclusion to installation evidence; performance and cleaning need their own studies.
2. Reconcile baseline revisions and assess affected checks — A consistent applicable baseline is needed to interpret the observations.
3. Resolve the identity discrepancy before relying on the match — A tag match does not resolve conflicting equipment identity.
4. As an observation whose acceptance criterion remains unresolved — The observed distance cannot create its own criterion; no universal 300 mm limit is taught.
5. Resolve the installed-gasket link before accepting the certificate — The missing item-to-evidence link must be resolved through an applicable controlled approach.
6. The cleaning process is effective for the intended residues — Matching material identity does not demonstrate cleaning effectiveness.
7. Record and investigate the connection discrepancy — A label does not correct an incorrect or unassessed physical connection.
8. Assessment of the change and applicable verification scope — The extension must be evaluated; previous qualification does not automatically cover a changed system.
9. +0.4 °C; within the fictional ±0.5 °C limit — Indication minus reference is +0.4 °C. This arithmetic is not a complete uncertainty or acceptance decision.
10. Resolve certificate applicability to the installed instrument — Identity, not just date and range, is necessary for applicability.
11. The validity of affected measurements and required follow-up — A new calibration record alone does not establish the validity of earlier evidence.
12. Validation of applicable functions and controls under the plan — Version identification does not demonstrate functional controls or complete Annex 11 compliance.
13. Verify applicability and obtain the required correct instructions — The handover needs instructions relevant to the installed model/configuration.
14. Applicable evidence plus justified configuration and impact review — Assess configuration, method, data and effects of transport/installation; retain required site checks.
15. Link the image to the inspected item and its verification record — An image needs a clear relationship to the relevant equipment and verification record.
16. Preserve the original and make an attributed, dated correction — Follow the approved procedure, preserve the original information and record a reason where appropriate.
17. Not checked, with the limitation and follow-up recorded — An incomplete check is distinct from a justified scope exclusion or a demonstrated pass.
18. Record the finding, assess impact and authorize the response — A discrepancy requires an evidence-based controlled response and defined verification.
19. Numerical coverage only; it does not authorize the next stage — 9/10 is 90%, but significance and activity-specific assessment govern the decision.
20. Documented no-significant-impact review and responsible authorization — Annex 15 §2.10 addresses conditional approval to proceed to the next qualification stage after a documented no-significant-impact assessment. Selecting a separately assessed dry OQ activity is a practical teaching recommendation, not an activity prescribed by that clause. The open finding remains open; responsible authorization is required.